| General Specifications |
| Measuring functions | 3-pole earth ground resistance, 2-pole AC resistance of a conductor,Interference voltage |
| Intrinsic error | Refers to the reference temperature range and is guaranteed for one year |
| Measuring rate | 2 measurements/second |
| Battery[1] | One 9 volt alkaline (LR61) |
| Battery condition | LO-BAT is displayed if voltage drops below 6.5 V |
| Voltages | 1. Between jacks H/C2 and E/C1: 250 Veff maximum (effective voltage) 2. Between jacks S/P2 and E/C1: 250 Veff maximum |
| Climatic class | VDE/VDI 3540 RZ (conforming to KWG as per DIN 40040, 4/87) |
| Temperature performance[2] | 1. Working: -10 °C to +50 °C (+14 °F to +122 °F) 2. Operating: 0 °C to +35 °C (+32 °F to +95 °F) 3. Storage: -20 °C to +60 °C (+68 °F to +140 °F) 4. Reference: +23 °C ± 2 °C (+73 °F ± 4 °F) |
| Temperature coefficient | ± 0.1% of range per degree Kelvin |
| Safety | IEC/EN 61010-1, 600 V CAT II, pollution degree 2 |
| Dimensions | 113mm x 54mm x 216mm (4.5 in x 2.1 in x 8.5 in), including holster |
| Weight | 850 g (1.9 lb.), including standard accessories, volume approximately 600 cm3 |
| Note: | [1]If the tester is not going to be used, or is being stored for a long period, remove the battery and store separately from the tester to avoid damage from battery leakage. |
| Note: | [2]The four temperature ranges for the tester exists to satisfy European Standards requirements; the instrument can be used over the full working temperature range by using the temperature coefficient to calculate accuracy at the ambient temperature of use. |